Series 240/241 Instruction Manual Table of Contents IM-24-A 0-1 Sierra Series 240 and 241 Innova-Mass™ Multi-Parameter Vortex Mass Flow Meters
Chapter 1 Introduction Series 240/241 Instruction Manual 1-4 IM-24-A Vortex Shedding Frequency Von Karman vortices form downstream of a shedder bar
Series 240/241 Instruction Manual Chapter 1 Introduction IM-24-A 1-5 Re = ρ V D μ St = f d V Flow Velocity Range To ensure trouble-free oper
Chapter 1 Introduction Series 240/241 Instruction Manual 1-6 IM-24-A As shown in Figure 1-3, Innova-Mass™ meters exhibit a constant Strouhal number
Series 240/241 Instruction Manual Chapter 1 Introduction IM-24-A 1-7 Pressure Measurement Innova-Mass Flow Meters incorporate a solid-state pressure
Chapter 1 Introduction Series 240/241 Instruction Manual 1-8 IM-24-A Flow Meter Electronics Innova-Mass Flow Meter electronics are available mounted
Series 240/241 Instruction Manual Chapter 2 Installation IM-24-A 2-1 Chapter 2 Installation Installation Overview Sierra’s Innova-Mass Vortex Flow
Chapter 2 Installation Series 240/241 Instruction Manual 2-2 IM-24-A Unobstructed Flow Requirements Select an installation site that will minimize p
Series 240/241 Instruction Manual Chapter 2 Installation IM-24-A 2-3 Series 240 In-Line Flow Meter Installation Install the Series 240 In-Line Flow
Chapter 2 Installation Series 240/241 Instruction Manual 2-4 IM-24-A Wafer-Style Flow Meter Installation Install the wafer-style meter between two c
Series 240/241 Instruction Manual Chapter 2 Installation IM-24-A 2-5 Flange-Style Flow Meter Installation Install the flange-style meter between two
Table of Contents Series 240/241 Instruction Manual 0-2 IM-24-A Customer Notice Unless you have specifically ordered Sierra’s optional O2 cleaning,
Chapter 2 Installation Series 240/241 Instruction Manual 2-6 IM-24-A Series 241 Insertion Flow Meter Installation Prepare the pipeline for installa
Series 240/241 Instruction Manual Chapter 2 Installation IM-24-A 2-7 Cold Tap Guidelines Refer to a standard code for all pipe tapping operations. T
Chapter 2 Installation Series 240/241 Instruction Manual 2-8 IM-24-A Hot Tap Guidelines Refer to a standard code for all pipe tapping operations. Th
Series 240/241 Instruction Manual Chapter 2 Installation IM-24-A 2-9 Flow Meter Insertion The sensor head must be properly positioned in the pipe.
Chapter 2 Installation Series 240/241 Instruction Manual 2-10 IM-24-A Installing Flow Meters with a Compression Connection* Use the following formul
Series 240/241 Instruction Manual Chapter 2 Installation IM-24-A 2-11 Insertion Procedure for Meters with a Compression Connection Figure 2-7. Fl
Chapter 2 Installation Series 240/241 Instruction Manual 2-12 IM-24-A Installing Flow Meters with a Packing Gland Connection* Use the formula below
Series 240/241 Instruction Manual Chapter 2 Installation IM-24-A 2-13 Insertion Procedure for Flow Meters with Permanent Insertion Tool Figure 2-9
Chapter 2 Installation Series 240/241 Instruction Manual 2-14 IM-24-A Insertion Procedure for Flow Meters with Removable Insertion Tool Figure 2-1
Series 240/241 Instruction Manual Chapter 2 Installation IM-24-A 2-15 7. Tighten the packing gland nuts to stop leakage around the stem. Do not tor
Series 240/241 Instruction Manual Table of Contents IM-24-A 0-3 Table of Contents Chapter 1 Introduction Innova-Mass™ Vortex Mass Flow Meters ...
Chapter 2 Installation Series 240/241 Instruction Manual 2-16 IM-24-A Insertion Procedure for Flow Meters with No Insertion Tool (Packing Gland Conn
Series 240/241 Instruction Manual Chapter 2 Installation IM-24-A 2-17 Adjusting Meter Orientation Depending on installation requirements, you may ne
Chapter 2 Installation Series 240/241 Instruction Manual 2-18 IM-24-A Enclosure Adjustment (Series 240 Only) Figure 2-13. Enclosure Viewing Adjust
Series 240/241 Instruction Manual Chapter 2 Installation IM-24-A 2-19 Wiring Connections The NEMA 4X enclosure contains an integral wiring compartme
Chapter 2 Installation Series 240/241 Instruction Manual 2-20 IM-24-A 4-20 mA Output Connections The standard Innova-Mass Flow Meter has a single 4-
Series 240/241 Instruction Manual Chapter 2 Installation IM-24-A 2-21 Figure 2-18. Non-Isolated 4–20 mA Output Using Meter Input Power Supply Fi
Chapter 2 Installation Series 240/241 Instruction Manual 2-22 IM-24-A Figure 2-20. Isolated Pulse Output with External Power Supply Figure 2-21
Series 240/241 Instruction Manual Chapter 2 Installation IM-24-A 2-23 Alarm Output Connections One alarm output (Alarm 1) is included on the standar
Chapter 2 Installation Series 240/241 Instruction Manual 2-24 IM-24-A Figure 2-24. Isolated Alarm Output Using Meter Provided Power Supply Remote E
Series 240/241 Instruction Manual Chapter 2 Installation IM-24-A 2-25 Optional Input Electronics Wiring The meter has two optional input wiring t
Table of Contents Series 240/241 Instruction Manual 0-4 IM-24-A Fluid Menu ...
Series M22/M23 Instruction Manual Chapter 3 Operation IM-22, IM-23 3-1 Chapter 3 Operating Instructions After installing the Innova-Mass Vortex Fl
Chapter 3 Operation Series M22/M23 Instruction Manual 3-2 IM-22, IM-23 Start-Up To begin flow meter operation: 1. Verify the flow meter is install
Series M22/M23 Instruction Manual Chapter 3 Operation IM-22, IM-23 3-3 Using the Setup Menus Programming the Flow Meter 1. Enter the Setup Menu
Chapter 3 Operation Series M22/M23 Instruction Manual 3-4 IM-22, IM-23 Output Menu
Series M22/M23 Instruction Manual Chapter 3 Operation IM-22, IM-23 3-5 Example for Setting an Output The following shows how to set Output 1 to meas
Chapter 3 Operation Series M22/M23 Instruction Manual 3-6 IM-22, IM-23 Display Menu PasswordENTERUsekeys to access menusDisplayMenuCycle Time(Sec)0
Series M22/M23 Instruction Manual Chapter 3 Operation IM-22, IM-23 3-7 Alarms Menu Example for Setting an Alarm The following shows how to set Ala
Chapter 3 Operation Series M22/M23 Instruction Manual 3-8 IM-22, IM-23 Totalizer #1 Menu PasswordENTERUsekeys to access menusTotalizerMenuTotaling
Series M22/M23 Instruction Manual Chapter 3 Operation IM-22, IM-23 3-9 Totalizer #2 Menu PasswordENTERUsekeys to access menusTotalizerMenuTotaling
Chapter 3 Operation Series 240/241 Instruction Manual 3-2 IM-24-A Energy Menu – For EMS Energy Meters Only Configuration: There are several pos
Series 240/241 Instruction Manual Table of Contents IM-24-A 0-5 Figures 1-1. In-Line Vortex Multi-Parameter Mass Flow Meter ... 1-3 1
Series 240/241 Instruction Manual Chapter 3 Operation IM-24-A 3-3 Fluid Menu Password ENTER Usekeys to access menusFluid Menu Flowing Fluid Liquids
Chapter 3 Operation Series 240/241 Instruction Manual 3-4 IM-24-A Units Menu Use the Units Menu to configure the flow meter with the desired units
Series 240/241 Instruction Manual Chapter 3 Operation IM-24-A 3-5 Time & Date Menu PasswordENTERUsekeys to access menusTime & DateMenuSet T
Chapter 3 Operation Series 240/241 Instruction Manual 3-6 IM-24-A Diagnostics Menu Use the Diagnostics Menu to simulate operation and review the
Series 240/241 Instruction Manual Chapter 3 Operation IM-24-A 3-7 Calibration Menu The Calibration Menu contains the calibration coefficients for
Chapter 3 Operation Series 240/241 Instruction Manual 3-8 IM-24-A Password Menu PasswordENTERUsekeys to access menusPasswordMenuSet Password1234ENT
Series 240/241 Instruction Manual Chapter 4 Troubleshooting & Repair IM-22/IM-23 4-1 Chapter 4 Troubleshooting and Repair Hidden Diagnostics M
Chapter 4 Troubleshooting & Repair Series 240/241 Instruction Manual 4-2 IM-22/IM-23 Column One Hidden Diagnostics Values • f = vortex sheddin
Series 240/241 Instruction Manual Chapter 4 Troubleshooting & Repair IM-22/IM-23 4-3 • Adj. Flilter = adjustable filter. Displays the filterin
Chapter 4 Troubleshooting & Repair Series 240/241 Instruction Manual 4-4 IM-22/IM-23 low alarm on the output. When High is selected it will giv
Table of Contents Series 240/241 Instruction Manual 0-6 IM-24-A Warnings and Cautions Warning! Agency approval for hazardous location installatio
Series 240/241 Instruction Manual Chapter 4 Troubleshooting & Repair IM-22/IM-23 4-5 o Slope = unique value for each set of electronics. o Int
Chapter 4 Troubleshooting & Repair Series 240/241 Instruction Manual 4-6 IM-22/IM-23 2. Mechanical installation may be incorrect. Verify the
Series 240/241 Instruction Manual Chapter 4 Troubleshooting & Repair IM-22/IM-23 4-7 vortex sensor from the analog board (the analog board is th
Chapter 4 Troubleshooting & Repair Series 240/241 Instruction Manual 4-8 IM-22/IM-23 1. For remote mounted electronics, carefully check all the
Series 240/241 Instruction Manual Chapter 4 Troubleshooting & Repair IM-22/IM-23 4-9 tronics stack in the enclosure. Loosen these screws (Note:
Chapter 4 Troubleshooting & Repair Series 240/241 Instruction Manual 4-10 IM-22/IM-23 Returning Equipment to the Factory Before returning any P
Series 240/241 Instruction Manual Appendix A Specifications IM-24-A A-1 Appendix A Product Specifications Accuracy Process Variables 240 Series I
Appendix A Specifications Series 240/241 Instruction Manual A-2 IM-24-A Nominal Pipe Size (in)Pressure0.50.7511.5234681.8 3 5 13 22 50 87 198 34718
Series 240/241 Instruction Manual Appendix A Specifications IM-24-A A-3 Nominal Pipe Size (in)Pressure0.50.7511.5234686.5 12 20 49 82 183 318 722 12
Appendix A Specifications Series 240/241 Instruction Manual A-4 IM-24-A 241 Pressure Ratings Probe Seal Process Connection Material Rating Orderin
Series 240/241 Instruction Manual Chapter 1 Introduction IM-24-A 1-1 Chapter 1 Introduction Innova-Mass™ Multi-Parameter Vortex Mass Flow Meters Th
Series 240/241 Instruction Manual Appendix A Specifications IM-24-A A-5 Pocess Fluid and Process Fluid: Ambient Temperature Standard temperature
Series 240/241 Instruction Manual Appendix B Glossary IM-24-A B-1 Er Appendix B Glossary A B C D A Cross sectional area. ACFM Actual Cubic Fe
Appendix B Glossary Series 240/241 Instruction Manual B-2 IM-24-A I J K L In-Line Flow Meter A flow meter which includes a short section of pipi
Series 240/241 Instruction Manual Appendix B Glossary IM-24-A B-3 Q R S T Q Flow rate, usually volumetric. Rangeability Highest measurable flo
Series 240/241 Instruction Manual Appendix C Fluid Calculations IM-24-A C-1 Appendix C Fluid Calculations Calculations for Steam T & P When “St
Appendix C Fluid Calculations Series 240/241 Instruction Manual C-2 IM-24-A Calculations for Gas (“Real Gas” and “Other Gas”) Use this formula to de
Series 240/241 Instruction Manual Appendix C Fluid Calculations IM-24-A C-3 Calculations for Liquid Use this formula to determine the settings for “
Series 240/241 Instruction Manual Appendix D HART Commands IM-24-A D-1 Appendix D HART Commands with DD Menu
Appendix D HART Commands Series 240/241 Instruction Manual D-2 IM-24-A Appendix D HART Commands without DD Menu Online Menu1 Device Setup2 PV3 PV
Chapter 1 Introduction Series 240/241 Instruction Manual 1-2 IM-24-A Note and Safety Information We use note, caution and warning statements through
Series 240/241 Instruction Manual Chapter 1 Introduction IM-24-A 1-3 How the Innova-Mass Vortex Mass Flow Meter Operates Figure 1-1. In-Line Vorte