Series 220/221 Instruction Manual Table of Contents IM-22 0-1 Sierra Series 220 and 221 Innova-Flo™ Vortex Flow Meters Instruction Manual
Chapter 1 Introduction Series 220/221 Instruction Manual 1-4 IM-22 Vortex Shedding Frequency Von Karman vortices form downstream of a shedder bar in
Series 220/221 Instruction Manual Chapter 1 Introduction IM-22 1-5 Re = ρ V D µ St = f d V Flow Velocity Range To ensure trouble-free operat
Chapter 1 Introduction Series 220/221 Instruction Manual 1-6 IM-22 As shown below, Innova-Flo™ Vortex Flow Meters exhibit a constant Strouhal number
Series 220/221 Instruction Manual Chapter 1 Introduction IM-22 1-7 Flow Meter Configurations Innova-Flo™ Vortex Flow Meters are available in two con
Chapter 1 Introduction Series 220/221 Instruction Manual 1-8 IM-22 The meter includes a local 2 x 16 character LCD display housed within the enclosu
Series 220/221 Instruction Manual Chapter 2 Installation Chapter 2 Installation Installation Overview Innova-Flo™ meter installations are simple and
Chapter 2 Installation Series 220/221 Instruction Manual Unobstructed Flow Requirements Select an installation site that will minimize possible disto
Series 220/221 Instruction Manual Chapter 2 Installation Series 220 In-Line Flow Meter Installation Install the Series 220 In-Line Flow Meter between
Chapter 2 Installation Series 220/221 Instruction Manual Wafer-Style Flow Meter Installation Install the wafer-style meter between two conventional p
Series 220/221 Instruction Manual Chapter 2 Installation Flange-Style Flow Meter Installation Install the flange-style meter between two conventional
Table of Contents Series 220/221 Instruction Manual 0-2 IM-22 Customer Notice Unless you have specifically ordered Sierra’s optional O2 cleaning, th
Chapter 2 Installation Series 220/221 Instruction Manual Series 221 Insertion Flow Meter Installation Prepare the pipeline for installation using ei
Series 220/221 Instruction Manual Chapter 2 Installation Cold Tap Guidelines Refer to a standard code for all pipe tapping operations. The following
Chapter 2 Installation Series 220/221 Instruction Manual Hot Tap Guidelines Warning! Hot tapping must be performed by a trained professional. US. re
Series 220/221 Instruction Manual Chapter 2 Installation Flow Meter Insertion The sensor head must be properly positioned in the pipe. For this reaso
Chapter 2 Installation Series 220/221 Instruction Manual Installing Flow Meters with a Compression Connection* Use the following formula to determine
Series 220/221 Instruction Manual Chapter 2 Installation Insertion Procedure for Meters with a Compression Connection Figure 2-7. Flow Meter with C
Chapter 2 Installation Series 220/221 Instruction Manual Installing Flow Meters with a Packing Gland Connection* Use the formula below to determine t
Series 220/221 Instruction Manual Chapter 2 Installation Insertion Procedure for Flow Meters with Permanent Insertion Tool Figure 2-9. Flow Meter w
Chapter 2 Installation Series 220/221 Instruction Manual Insertion Procedure for Flow Meters with Removable Insertion Tool Figure 2-10. Flow Meter
Series 220/221 Instruction Manual Chapter 2 Installation 7. Tighten the packing gland nuts to stop leakage around the stem. Do not torque over 20 ft
Series 220/221 Instruction Manual Table of Contents IM-22 0-3 Table of Contents Chapter 1 Introduction Innova-Flo™ Vortex Flow Meters...
Chapter 2 Installation Series 220/221 Instruction Manual Insertion Procedure for Flow Meters with No Insertion Tool (Packing Gland Connection) 1. C
Series 220/221 Instruction Manual Chapter 2 Installation Adjusting Meter Orientation Depending on installation requirements, you may need to adjust t
Chapter 2 Installation Series 220/221 Instruction Manual Enclosure Adjustment (Series 220 Only) Figure 2-13. Enclosure Viewing Adjustment To avoid
Series 220/221 Instruction Manual Chapter 2 Installation Wiring Connections Warning! To avoid potential electric shock, follow National Electric Cod
Chapter 2 Installation Series 220/221 Instruction Manual 4-20 mA Output Connections The Innova-Flo meter has a single 4-20 mA loop. The 4-20 mA loop
Series 220/221 Instruction Manual Chapter 2 Installation Pulse Output Connections The pulse output is used for a remote counter. When the preset volu
Chapter 2 Installation Series 220/221 Instruction Manual 2-22 IM-22
Series 220/221 Instruction Manual Chapter 2 Installation Remote Electronics Wiring The remote electronics enclosure should be mounted in a convenient
Series 220/221 Instruction Manual Chapter 3 Operation Chapter 3 Operating Instructions After installing the Innova-Flo Vortex Meter, you are ready t
Chapter 3 Operation Series 220/221 Instruction Manual Start-Up To begin flow meter operation: Note Starting the flow meter or pressing EXIT will al
Table of Contents Series 220/221 Instruction Manual 0-4 IM-22 Chapter 4 HART Communications Wiring ...
Series 220/221 Instruction Manual Chapter 3 Operation Using the Setup Menus OutputMenuDisplayMenuTotalizerMenuUnitsMenuDiagnosticsMenuCalibrationMen
Chapter 3 Operation Series 220/221 Instruction Manual Output Menu PasswordENTERUsekeys to access menusOutputMenu4-20mA Output 1More >ENTERRun Mo
Series 220/221 Instruction Manual Chapter 3 Operation Display Menu PasswordENTERUsekeys to access menusDisplayMenuCycle Time(sec)0ENTERRun ModeNumb
Chapter 3 Operation Series 220/221 Instruction Manual Totalizer Menu PasswordENTERUsekeys to access menusTotalizerMenuTotalingInactiveMassVolumeENTE
Series 220/221 Instruction Manual Chapter 3 Operation Units Menu PasswordENTERUsekeys to access menusUnitsMenuENTERRun Modelb = poundsSton = 2000 p
Chapter 3 Operation Series 220/221 Instruction Manual Diagnostics Menu PasswordENTERUsekeys to access menusDiagnosticsMenuSim Vor FreqxxxENTERRun M
Series 220/221 Instruction Manual Chapter 3 Operation Calibration Menu PasswordENTERUsekeys to access menusCalibrationMenuMeter Sizeor Pipe IDENTERR
Chapter 3 Operation Series 220/221 Instruction Manual Password Menu PasswordENTERUsekeys to access menusPasswordMenuSet Password1234ENTERRun Mode
Series 220/221 Instruction Manual Chapter 4 HART Communications Chapter 4 HART Communications Wiring The diagram below detai
Chapter 4 HART Communications Series 220/221 Instruction Manual HART Menus Online Menu1 Device Setup2 PV3 PV AO4 PV LRV5 URV1 Process Variables2 Diag
Series 220/221 Instruction Manual Table of Contents IM-22 0-5 Warnings and Cautions Warning! Agency approval for hazardous location installations
Series 220/221 Instruction Manual Chapter 4 HART Communications Fast Key Sequence Use password 16363. Sequence Description Access Notes 1,1,1 Snsr
Chapter 4 HART Communications Series 220/221 Instruction Manual Sequence Description Access Notes 1,4,3,1,3,1 4 mA View Loop test, fix analog o
Series 220/221 Instruction Manual Chapter 5 Troubleshooting & Repair Chapter 5 Troubleshooting and Repair f fiG
Chapter 5 Troubleshooting & Repair Series 220/221 Instruction Manual Column One Hidden Diagnostics Values • f = vortex shedding frequency (Hz)
Series 220/221 Instruction Manual Chapter 5 Troubleshooting & Repair • Pulse Out Queue = Pulse output queue. This value will ac-cumulate if the
Chapter 5 Troubleshooting & Repair Series 220/221 Instruction Manual agnostics) and then actuate the enter key twice. This action will cause the
Series 220/221 Instruction Manual Chapter 5 Troubleshooting & Repair tice, the Ck value controls the adaptive filter, fi, setting. During flow,
Chapter 5 Troubleshooting & Repair Series 220/221 Instruction Manual Electronics Assembly Replacement (All Meters) The electronics boards are el
Series 220/221 Instruction Manual Chapter 5 Troubleshooting & Repair Returning Equipment to the Factory Before returning any Innova-Flo meter
Chapter 5 Troubleshooting & Repair Series 220/221 Instruction Manual 5-8 IM-22
Table of Contents Series 220/221 Instruction Manual 0-6 IM-22
Series 220/221 Instruction Manual Appendix A Specifications IM-22 A-1 Appendix A Product Specifications Accuracy Process 220 Series In-Line Mete
Appendix A Specifications Series 220/221 Instruction Manual A-2 IM-22 Water Minimum and Maximum Flow Rates ½-inch ¾-inch 1-inch 1.5-inch 2-inch
Series 220/221 Instruction Manual Appendix A Specifications IM-22 A-3 Power Requirements Loop powered, 12 to 36 VDC. Output Signals Analog: fiel
Appendix A Specifications Series 220/221 Instruction Manual A-4 IM-22
Series 220/221 Instruction Manual Appendix B Glossary Appendix B Glossary A B C D A Cross sectional area. acfm Actual cubic feet per minute (v
Appendix B Glossary Series 220/221 Instruction Manual I J K L In-Line Flow Meter A flow meter which includes a short section of piping which is p
Series 220/221 Instruction Manual Appendix B Glossary Q R S T Q Flow rate, usually volumetric. Rangeability Highest measurable flow rate divide
Appendix B Glossary Series 220/221 Instruction Manual B-4 IM-22
Series 220/221 Instruction Manual Chapter 1 Introduction IM-22 1-1 Chapter 1 Introduction Innova-Flo™ Vortex Flow Meters The Sierra Instruments’ Se
Chapter 1 Introduction Series 220/221 Instruction Manual 1-2 IM-22 Note and Safety Information We use note, caution and warning statements throughou
Series 220/221 Instruction Manual Chapter 1 Introduction IM-22 1-3 How the Innova-Flo Vortex Flow Meter Operates Figure 1-1. Series 220 In-Line Vo